Open Access
ARTICLE
Microperforations of surgical gloves in urology: minimally invasive versus open surgeries
James Buchanan Brady Urological Institute, Department of Urology, Johns Hopkins Medical Institutions, Baltimore, Maryland, USA
Address correspondences to Dr. Mohamad E. Allaf, James
Buchanan Brady Urological Institute, 600 N. Wolfe Street,
Park 223, Baltimore, MD 21230 USA
Canadian Journal of Urology 2011, 18(2), 5615-5618.
Abstract
Introduction: Surgical glove integrity is important in preventing wound infections and reducing patient mortality. Rates of perforations have been studied in many surgical subspecialties, but glove perforations specific to urology have not been investigated previously. This study aims to determine the incidence of glove perforations during urological surgeries and to investigate differences between open, laparoscopic, and endoscopic procedures.Materials and methods: A total of 180 gloves were collected from various urological procedures performed at our institution: 59 from endoscopic, 72 from laparoscopic, and 49 from open cases. The gloves were tested for defects by both the water load test and electrical conductance testing. The frequency of defects for each type of procedure along with the length of wear, surgeon experience, and glove brand was analyzed.
Results: Glove defects were detected in 29% of all cases. Microperforations encompassed the majority of the glove defects (23.3%) and were detected in 15.2%, 25.0 %, and 30.6% of the endoscopy, laparoscopic and open surgical cases, respectively. The frequency of perforations noted in the minimally invasive procedures was significantly different across the groups (p < 0.01). There was no statistical significant correlation between glove defects and operation time, surgeon experience, and glove brand.
Conclusions: The rates of glove perforation (29%) in urological procedures were higher than expected. Given the high rates of glove perforations found, double gloving in urological surgeries may offer a solution to the increased risk for cross contamination from microscopic perforations.
Keywords
Cite This Article
Copyright © 2011 The Author(s). Published by Tech Science Press.This work is licensed under a Creative Commons Attribution 4.0 International License , which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


Submit a Paper
Propose a Special lssue
Download PDF
Downloads
Citation Tools