    ARTICLE

    Nit-Occlud Lê VSD coil versus Duct Occluders for percutaneous perimembranous ventricular septal defect closure

    Hieu Lan Nguyen1,2, Quang Tan Phan3,4, Linh Huynh Dinh2, Hieu Ba Tran2, Hoyoun Won4, Julian Johny Thottian4,5, Dung Doan Duc1, Truong Nguyen Quang3, Sang Wook Kim4

    Congenital Heart Disease, Vol.13, No.4, pp. 584-593, 2018, DOI:10.1111/chd.12613

    Abstract Objective: To evaluate the efficacy and safety of Nit-Occlud Le VSD Coil versus Duct Occluders ^ for percutaneous closure of perimembranous ventricular septal defect (pmVSD).
    Introduction: VSD closure using conventional pmVSD occluders has been largely abandoned because of an unacceptable high rate of complete heart block (CHB). The advantages of Duct Occluders and VSD Coil are supposed to reduce the drawbacks of previous devices, especially CHB complications.
    Method: Patients underwent percutaneous pmVSD closure were divided into Coil group (using VSD Coil, n = 71) and DO group (using Duct Occluders, n = 315). Patient demographics, clinical presentations, echocardiography measurements, procedure… More >